Health Products

Teleflex Stylet, Endotracheal Kit and Flexi-Set

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Teleflex Stylet, Endotracheal Kit and Flexi-Set
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
KME21E1286Flexi-Set504580
KME21D1430
KME21E1290
KME21E1287
KME21E1232
KME21E1296
Slick Set Stylet - Uncuffed170080
170090
170085
170070
KME21E0660
KME21E0021
KME21E0607
KME21E0002
KME21F1006
KME21D2020
KME21D2929
KME21E0288
KME21F0981
KME21F0946
KME21F0741
KME21D1946
KME21E0498
KME21F0930
Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff112082055
112082060
112082070
112082065
112082075
112082080
112082085

Issue

Teleflex has initiated a voluntary field safety corrective action due to reports of pilot balloon non-inflation or cuff non-deflation. In the event of pilot balloon non-inflation, there is a potential risk of injury to the patient as the pilot balloon no longer reflects the condition of the cuff, which may lead to over-inflation of the cuff. The risk persists in situ, therefore clinicians should evaluate the risk benefit to individual patients of extubation and reintubation. In the event of cuff non-deflation, there is the potential risk of injury if the device is removed from the patient with the cuff still inflated. If extubation and removal of the endotracheal tube is planned and the cuff cannot be deflated via the bespak valve, then the inflation line should be cut in order to facilitate passive deflation of the cuff.

Recall start date: December 9, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime