Health Products

Alinity M System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Alinity M System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
All lots.08N53-002Alinity M System

Issue

Abbott has identified three potential issues which may have impacted the amplification detection unit(s) (ADU) installed on your Alinity M system. 1. systems which had their ADU's serviced may have an incorrect boardvalue (calibration data file) referenced causing a misconfiguration of the calibration. 2. inability of maintenance and diagnostics 2300 (amp detect optical calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. under specific conditions during the adu manufacturing process, an incorrect calibration may potentially have been configured.

Recall start date: December 13, 2022