Health Products

Epoc Blood Analysis System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry, Health products - Medical devices - General hospital and personal use
Product: Epoc Blood Analysis System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
EPOC HOST 2, CA
HOST 2, EPOC, REFURBISHED
HH-1009-CA-00
HH-1R09-CA-00
HH-1009-00-00
Epoc Blood Analysis System
epoc NXS Host - 5553CA11413506Epoc Blood Analysis System - Computing Device With Custom Software

Issue

Siemens Healthcare Diagnostics Inc. has confirmed an issue through complaint investigation. The issue has the potential to affect the recovery of po2 in quality control (qc) at elevation with barometric pressures <730 mmhg on epoc sensor configuration 41.1, epoc host 2 sw version 3.37.3 and epoc nxs host sw version 4.10.6, which were released on 17 October 2022.

Recall start date: December 19, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime