Health ProductsType I recall

Trilogy Ventilators

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Trilogy Ventilators
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Not applicable.CA1032800
CA1032800B
Trilogy 200, Canada
Not applicable.CA1054096
CA1054096B
Trilogy 100 Ventilator-Canada

Issue

The replacement silicone sound abatement foam installed into the Trilogy 100 and Trilogy 200 devices identified in this letter may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm. Additionally, Philips Respironics has observed residual pe-pur sound abatement foam in some devices that were returned to the field. these cases were limited but further exposure to pe-pur sound abatement foam should be avoided.

Recall start date: December 16, 2022