· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Orise Gel Twin Syringe Kit With Interject | Not applicable |
| All lots. | Orise Proknife Procedure Kit | M00519400 M00519380 M00519390 |
| All lots. | Orise Gel Syringe Single Pack | M00519231 M00519210 M00519220 M00519221 M00519230 M00519211 |
| All lots. | Orise Gel Syringe Twin Pack | M00519201 M00519200 |
Boston Scientific (BSC) initiated a field action in October 2022, communicating that remnant orise gel post-procedure can cause a foreign body reaction, appearing as mass formations and submucosal distortions. Since this time, BSC has become aware of additional events leading to unnecessary medical and surgical intervention. As a result, BSC now believes the occurrence is potentially higher than anticipated, therefore, this product is being removed from the market.
Recall start date: December 15, 2022