· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 170369 170382 | Artis Icono Floor | 11327700 |
| 180415 180324 | Artis Icono Biplane | 11327600 |
| 164101 | Artis Pheno | 10849000 |
During transversal table movement, it could happen that the rolling bearing which is part of the linear guide rail and enables the transversal table movement may get dislocated. The rolling bearing may fall out of the linear guide rail and thus the mechanical connection between table top and table base may get lost.
Recall start date: December 19, 2022