Health Products

Magnetom Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Magnetom Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
211019Magnetom Avanto Fit11516216
More than 10 numbers, contact manufacturer.Magnetom Aera10432914
166026
166139
166247
66071
166208
66108
Magnetom Prisma10849582
189602
189915
189956
189700
190027
Magnetom Altea11344915
More than 10 numbers, contact manufacturer.Magnetom Sola11291455
202567
202589
202562
202570
Magnetom Sola Fit11410482
More than 10 numbers, contact manufacturer.Magnetom Vida11060815
196408
196406
196427
Magnetom Lumina11344916
45186
45281
45446
45166
45087
145249
45768
145034
45934
45961
45370
145335
45447
Magnetom Skyra10432915

Issue

Sporadically, the distance measurement values displayed in MRI images on third-party postprocessing systems might be wrong if: - enhanced MRI images with rectangular FOV (field of view) are created at the MR scanner, and - in Syngo.via or MR view&go, the user rotates the images to display them in any orientation and saves them, and - afterwards, the user exports the images to another Dicom node using MR export rules (interoperability mode) the distance measurement values on newly created series might be displayed incorrectly on systems that receive these new images.

Recall start date: December 21, 2022