Health Products

Cardiosave System - Hybrid & Rescue Intra-Aortic Balloon Pump

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Cardiosave System - Hybrid & Rescue Intra-Aortic Balloon Pump
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Cardiosave System - Hybrid & Rescue Intra-Aortic Balloon Pump0998-00-0800-XX

Issue

Datascope/Getinge has received complaints reporting unexpected shutdown of the Cardiosave IABP while providing therapy in very rare instances. An internal investigation of the complaints determined an unexpected shutdown may be due to blood entering into the Cardiosave IABP when therapy is provided with a perforated intra-aortic balloon catheter.

Recall start date: December 21, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime