Health Products

Ortho Vision Analyzers

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Ortho Vision Analyzers
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Ortho Vision Analyzer6904577
More than 10 numbers, contact manufacturer.Ortho Vision Max Analyzer6904576

Issue

Ortho's investigation discovered that the self-service customer procedures guide, both onboard and offboard, contains 2 errors related to the probe installation/removal process (chapter 4).

1. The self-service customer procedure guide contains a note indicating the probe can be replaced without using the software maintenance task. If it is performed this way, the instructions do not instruct the operator to condition the probe (daily maintenance) after replacement, which is a required post requisite activity.

2. The self-service customer procedure guide contains a note indicating the software will initiate the pump test, after probe replacement is complete. This is incorrect, since upon completion of the probe replacement maintenance task, the software marks the pipetting volume test as pending rather than the pump test. Note: if the user follows the software's on-screen instructions, daily maintenance is automatically performed during the probe replacement task and the operator is notified of the pending pipetting volume test, which if completed, mitigates this issue.

To date, no customer complaints have been recorded regarding this issue.

Recall start date: December 21, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime