· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| All lots. | Biostop G Bioresorbable Cement Restrictor | 546314000 546308000 546316000 546320000 546310000 546312000 546318000 |
All lots of Biostop G Bioresorbable Cement Restrictor are being removed as a precautionary measure because recent in vitro testing of endotoxin levels from a sample restrictor fully dissolved in solution were calculated at >20,000 endotoxin units (eu)/device over 24 hours, based on restrictor size. This exceeds the recommended value of 20 eu/device over 1 hour as referenced in the current FDA regulatory guidance.
Recall start date: December 21, 2022