Health Products

Biostop G Bioresorbable Cement Restrictor

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Biostop G Bioresorbable Cement Restrictor
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
All lots.Biostop G Bioresorbable Cement Restrictor546314000
546308000
546316000
546320000
546310000
546312000
546318000

Issue

All lots of Biostop G Bioresorbable Cement Restrictor are being removed as a precautionary measure because recent in vitro testing of endotoxin levels from a sample restrictor fully dissolved in solution were calculated at >20,000 endotoxin units (eu)/device over 24 hours, based on restrictor size. This exceeds the recommended value of 20 eu/device over 1 hour as referenced in the current FDA regulatory guidance.

Recall start date: December 21, 2022

Get told about the next health product recall.
Get health product recall alerts