Somatom X.Cite

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Somatom X.Cite
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
123296Somatom X.Cite11330001

Issue

In the Syngo.ct va40 and Syngo.ct va50 myneedle guide application, the order of the orientation labels "head", "center", and "feet" in the axial thick slice segments are erroneously swapped in the following two cases: • if the recon parameter "image order" is changed by the user from the default setting "head to feet" to "feet to head", or • if the recon parameter "mirroring" is changed by the user from the default setting "none" to any other value, this influences the parameter "image order", which is switched from the default setting "head to feet" to "feet to head" so that above-mentioned condition applies.