· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | HLS Set Advanced 7.0 With Bioline Coating | 701069065 BEQ-HLS 7050-CA |
| More than 10 numbers, contact manufacturer. | Permanent Life Support Set | BE-PLS 2051 BE-PLS 2050 |
| More than 10 numbers, contact manufacturer. | HLS Set Advanced 5.0 With Bioline Coating | 701069068 BEQ-HLS 5050-CA |
Potentially compromised sterile barrier. The conformity of the products was questioned by the German health authority due to packaging tests not adequately performed. Single sterilized and transport conditioned according to ASTM d4169-22 was conducted. However, the test samples were not double sterilized to cover an assumed worst condition of sterilization impact. These required tests are currently under way, however these test results will not be available until earliest in April 2023.
Recall start date: January 9, 2023