· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 61491501 64100801 63986301 64241701 | Elecsys® Troponin T Hs | 08469717190 08469873190 |
| 59638102 56661101 59638101 63480901 56661102 56661103 | Elecsys® Troponin T Hs Stat | 08469814190 |
During internal studies with the Elecsys® Troponin T hs (high sensitive) and Elecsys® Troponin T hs Stat assays, it was confirmed that for certain k2/k3 EDTA primary tubes, TnT hs results are elevated compared to serum samples when measured from the primary tube after processing the sample according to the tube manufacturers' instructions. This issue does not affect lithium heparin plasma or serum sample tubes.
Recall start date: January 16, 2023