Access 2 Immunoassay Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Access 2 Immunoassay Analyzer
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Access 2 Immunoassay Analyzer81600N

Issue

Beckman Coulter has determined that the Access 2 wash arm lead screw nut band is susceptible to breakage. An Access 2 Immunoassay system wash arm may not move as intended if the nut band is broken. A broken nut band may increase the risk of erroneous results.

Recall start date: January 18, 2023