Health Products

Anesthesia Set Flex, Latex Free, Us

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Anesthesia Set Flex, Latex Free, Us
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
All lots.MP00303-XXAnesthesia Set Flex, Latex Free, Us

Issue

In the course of internal quality controls, Draeger has noticed that some of their flex ventilation hoses have contaminations. Due to a deviation in the manufacturing process, components of the hose material remained as particles inside the hoses. Draeger's analyses have shown that particles are not reliably detected by visual inspection prior to use. So far, they have not received any customer feedback on the mentioned contamination. However, they cannot rule out the possibility that they have shipped affected hoses.

If the contaminated flex ventilation hoses are used without filters on the patient side, it cannot be ruled out that particles enter the airway of the affected patients and cause coughing, irritation or toxicological effects.

Recall start date: January 24, 2023