Health Products

Quanta Flash Gbm Calibrators

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Quanta Flash Gbm Calibrators
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
223636Quanta Flash Gbm Calibrators701141

Issue

As a result of an investigation of the above-mentioned lot, we have confirmed that the calibrator values assigned to this lot are erroneous. These erroneous value assignments would lead to an increase in patient values and could therefore lead to overestimation of patient results, and potentially a false positive result.

Recall start date: January 30, 2023