Health Products

Freestyle Libre Sensor Kits

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry, Health products - Medical devices - Gastroenterology and urology
Product: Freestyle Libre Sensor Kits
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
KTP00505871534-01Freestyle Libre Flash Glucose Monitoring System - Sensor Kit
KTP00508071993-01Freestyle Libre 2 Sensor Kit

Issue

It was identified that two (2) Freestyle Libre sensor reel lots manufactured by Abbott Diabetes Care (ADC) failed retain testing. As a result of the investigation and testing of retains, the sensors were found to give clinically significant readings results; sensors may provide erroneous high sensor glucose results. These sensor reel lots were manufactured into six (6) finished goods lots (Freestyle Libre and Freestyle Libre 2 sensors) and distributed to the field. The lots have been distributed to Japan, Canada, Saudi Arabia, and United Kingdom. In Canada, this issue affects a subset of Freestyle Libre® 2 sensors with the lot number KTP005080 and Freestyle Libre® sensors with the lot number KTP005058. If you have these lot numbers, please visit www.freestyleconfirm.com to see if your sensor is impacted.

Recall start date: February 3, 2023