· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| KTP005058 | 71534-01 | Freestyle Libre Flash Glucose Monitoring System - Sensor Kit |
| KTP005080 | 71993-01 | Freestyle Libre 2 Sensor Kit |
It was identified that two (2) Freestyle Libre sensor reel lots manufactured by Abbott Diabetes Care (ADC) failed retain testing. As a result of the investigation and testing of retains, the sensors were found to give clinically significant readings results; sensors may provide erroneous high sensor glucose results. These sensor reel lots were manufactured into six (6) finished goods lots (Freestyle Libre and Freestyle Libre 2 sensors) and distributed to the field. The lots have been distributed to Japan, Canada, Saudi Arabia, and United Kingdom. In Canada, this issue affects a subset of Freestyle Libre® 2 sensors with the lot number KTP005080 and Freestyle Libre® sensors with the lot number KTP005058. If you have these lot numbers, please visit www.freestyleconfirm.com to see if your sensor is impacted.
Recall start date: February 3, 2023