Health Products

Delfia Inducer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Delfia Inducer
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
742762
742802
3027-0010Delfia Inducer

Issue

On Delfia Inducer bottles (2 kit lots), the label is placed too low preventing the Delfia Xpress instrument level sensor to detect the Delfia inducer liquid level. Because of that the instrument does not give an 'inducer bottle soon empty' warning at level where there is 4 ml Delfia inducer left (30 tests + 1 ml dead volume). This may cause the Delfia inducer to run out and this in turn may cause failed or false results.

Recall start date: February 7, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime