Health Products

Cardiomems Patient and Hospital Electronics System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Cardiomems Patient and Hospital Electronics System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Cardiomems Patient Electronics SystemCM1100
All lots.Cardiomems Hospital Electronics SystemCM3000

Issue

Abbott has identified that when the Cardiomems Patient Electronics Systems (model CM1100) and Cardiomems Hospital Electronics Systems (HES) (model CM3000) are used to take a reading with the pulmonary artery (pa) sensor, the radiofrequency (RF) emissions at certain frequencies are higher than levels listed in the instructions for use (IFU).

Recall start date: February 7, 2023