| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| VD10U VD10P | Uroskop Omnia | 10094910 |
| VF10F_HF03 VE10Y VF11C VF10F VF11E VF10C VF11C_HF01 | Luminos Drf Max | 10762471 |
| Not applicable | Luminos Agile Max | 10762472 |
| VF11G VF10F VE10Z VF11C_HF01 | Uroskop Omnia Max | 10762473 |
| VD10U VD10T | Axiom Luminos Drf | 10094200 |
| VF11C VF11G | Luminos Lotus Max | 11574100 |
| VC10Q | Luminos Agile | 10502200 |
Results of post market surveillance revealed that under certain circumstances the room configuration parameters of the system are set to default values. If these values are larger than the actual room dimensions, a collision with the ceiling or wall cannot be ruled out during system movement. Siemens Healthineers has not become aware of any injuries caused by this issue.
Recall start date: February 16, 2023