· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| H5802983 | 3604024 | Infinity Occipitocervical Upper Thoracic System Multi Axial Screw |
It was identified by Medtronic that an incorrect type of titanium was used for the manufacturing of subcomponents of multi-axial screws that are for use as part of the Infinity oct system. This incorrect material reduces the clamping strength of the multi-axial screw, which may lead to a loss of fixation with the infinity system construct.
Recall start date: February 16, 2023