Health Products

Da Vinci Surgical Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology, Health products - Medical devices - General hospital and personal use
Product: Da Vinci Surgical Systems
Issue: Medical devices - Performance
What to do: CAPA 552131 started to address issue. Customers provided with further instructions in response.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Contact manufacturer.Da Vinci X Surgical System Is4200PS4200
Contact manufacturerDa Vinci Xi Surgical System Is4000IS4000

Issue

An increase in reports in which a loose instrument carriage is observed on the universal side manipulator (usm) or the instrument arm of the Da Vinci X and Xi systems. The loose instrument carriage is an outcome of a linear rail component being out of specification. The loose instrument carriage makes it possible for the carriage to wobble along the linear rail in response to force applied to the carriage or to the shaft of an installed instrument.

Recall start date: February 23, 2023