· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | BD Bodyguard Microset | 120-160XYSK ONC00005 100-160XSMG90EK 120-112XPEFKY EPI00011-NRF PCA00001 TPN00005 100-163XE90SK 100-163XESK 100-163XESVK 100-184XSYK 120-160XCSEK 120-160XSFMK EPI00003 ONC00004UV 120-003XS2YV PCA00004 100-163XNKS 120-160XCSK EPI00012-NRF 100-184XSK TPN00004 101-163XE90SK 120-112XSFK 120-112XSFVK 120-160X90SK ONC00006 PCA00002 |
This recall is being initiated to notify customers that BD has identified internally through a product review that the BD Bodyguard™ Microsets do not currently have the evidence to support compliance with ethylene oxide (eo) residual level requirements (per ISO 10993-7:2008/amd. 1:2019) for use in special populations, specifically children, infants, neonates and pre-mature neonates. Therefore, these devices are currently being limited in use with adults until the compliance evidence is gathered.
Recall start date: February 13, 2023