Durasul Modular Head

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Durasul Modular Head
Issue: Medical devices - Device compatibility, Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Durasul Modular Head01.01012.384

Issue

Zimmer Biomet is conducting a medical device correction to update the compatibility matrix as referred to in the instructions for use (IFU) for the CoCr Femoral Head XS. The update is to remove the compatibility with the Epsilon Durasul Constrained Acetabular liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009). When used in combination with the Epsilon Durasul Constrained Acetabular liner, the CoCr Femoral Head XS has a range of motion of 76 degrees. The reduced range of motion could potentially result in the health risks identified. there have been no complaints for this issue.

Recall start date: March 3, 2023