Health ProductsType I recall

Infinity® Central Station

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular, Health products - Medical devices - General hospital and personal use
Product: Infinity® Central Station
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Infinity® M300+MS33648
MS33659
MS33950
All lots.Infinity® M300MS 25755
MS18501
MS 26076
All lots.Infinity® Centralstation WideMS26800
All lots.Infinity® Central StationMS 18500
MS26800

Issue

During our global market surveillance activities, we have become aware of cases in which the software on the Infinity® Central Station drops peaks on narrow waveforms causing the Infinity® M300 to fail the requirements of the standard IEC 60601-2-27.

Recall start date: March 9, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime