Health Products

Trilogy EV300

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Trilogy EV300
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Trilogy EV300CA2200X12B

Issue

Philips Respironics has discovered that accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FioO2 sensor may indicate a value higher than the device is actually delivering. This may vary based on the patient's lung capacity, lung resistance, use of a particulate filter, or circuit configuration. In the worst case, this may lead to under delivery of oxygen.

Recall start date: March 3, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime