Mridium MRI Infusion System - Syringe Adapter Set

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Mridium MRI Infusion System - Syringe Adapter Set
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
22D58N
22A59N
22A03N
22I17N
22J46N
22B54N
Mridium MRI Infusion System - Syringe Adapter Set1057

Issue

Finished 1057 sets made with a occluded air pathway will trigger the inlet occlusion alarm of the pump and stop delivery of medication until the inlet occlusion is resolved by replacement of the 1057 IV set. The pump detects the inlet occlusion functions as specified, the occluded air venting pathway is not per specification for the 1057 IV set. This can lead to a hazard situation of a delay of therapy and possible under-dosing which could result in serious adverse health events.

Recall start date: February 27, 2023