Health Products

Philips Azurion System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Philips Azurion System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Azurion 7 M20722 224
722 079
More than 10 numbers, contact manufacturer.Azurion 7 M12722 223
722 078
151
188
Azurion 7 B12722 067
149
296
Azurion 7 B20722 226
722 068

Issue

Philips Azurion System may unexpectedly lose X-ray functionality. Due to a software issue, a mechanism that is present in the system to manage the number and size of log trace files does not function properly. Without this mechanism, the log trace files created by the system (e.g., at start, during use) may occupy the full disk capacity of the Philips Azurion system. When the full disk capacity is reached, X-ray functionality will cease to be available without an advance warning to the user.

Recall start date: March 23, 2023