· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Azurion 7 M20 | 722 224 722 079 |
| More than 10 numbers, contact manufacturer. | Azurion 7 M12 | 722 223 722 078 |
| 151 188 | Azurion 7 B12 | 722 067 |
| 149 296 | Azurion 7 B20 | 722 226 722 068 |
Philips Azurion System may unexpectedly lose X-ray functionality. Due to a software issue, a mechanism that is present in the system to manage the number and size of log trace files does not function properly. Without this mechanism, the log trace files created by the system (e.g., at start, during use) may occupy the full disk capacity of the Philips Azurion system. When the full disk capacity is reached, X-ray functionality will cease to be available without an advance warning to the user.
Recall start date: March 23, 2023