· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Ventstar Anesthesia (N) 180 | MP00333 |
| More than 10 numbers, contact manufacturer. | Ventstar Anesthesia Wt (P)180 | MP00374 |
| More than 10 numbers, contact manufacturer. | Ventstar Coax (P) 150 | MP00379 |
| More than 10 numbers, contact manufacturer. | Ventstar Watertrap (P) 180 W/O Ll | MP00362 |
| More than 10 numbers, contact manufacturer. | Ventstar Basic (P) 250 | MP00352 |
| More than 10 numbers, contact manufacturer. | Ventstar Basic (P) 180 | MP00351 |
| More than 10 numbers, contact manufacturer. | Ventstar Bag Set | MP00384-XX MP00383-XX |
| More than 10 numbers, contact manufacturer. | Ventstar Basic (N) 180 | MP00353 |
| More than 10 numbers, contact manufacturer. | Seattle Pap Plus | MP02260 |
During global market surveillance activities, manufacturer became aware of cases in which glued connections of the breathing circuits became loose either before or during the ventilation process resulting in partial or complete detachment of components such as watertrap, y-piece or hose connector. There were no reports of any adverse impact on the health of the patients involved. Investigations have identified that the adhesive joint did not conform to specification due to a manufacturing error. If system components become loose and/or detach from the hose during use, ventilation of the patient will be restricted.
Recall start date: March 22, 2023