Health Products

Integre Pro

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ophthalmology
Product: Integre Pro
Issue: Medical devices - Performance
What to do: Contact CLARION MEDICAL TECHNOLOGIES INC
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
IPG0542
SN: IPG0542
Integre ProLP5532

Issue

Ellex was recently made aware of a potential safety issue impacting a small number of Integre Pro/Integre Pro Scan lasers . This was reported as a single event during servicing with no injury; when using a laser-indirect-ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may exit through the slit lamp output port (SDS) and get emitted from the objective lens when the laser is fired. The cause was traced to an incorrect assembly of a sub-module in the laser cavity system.

Recall start date: March 13, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime