Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) Assay

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) Assay
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Beckman Coulter Au5400 Immunochemistry System - Bicarbonate (CO2)AssayOSR6237
OSR6137
All lots.Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) AssayOSR6237
OSR6137
All lots.Beckman Coulter Au2700 Immunochemistry System - Bicarbonate(CO2) AssayOSR6237
OSR6137

Issue

Beckman Coulter has confirmed a premature decrease in calibration ODs for several lots of bicarbonate OSR6X37. It is possible that all bicarbonate lots may also be impacted by this issue, therefore all lots are subject to this field action.

The abnormal decrease in calibrator ODs may result in calibration failures and/or OC failures. the risk of generating falsely elevated, clinically impactful patient test results is remote.

Recall start date: April 3, 2023