· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Duralife® Autoclavable Silicone Extension Tube | 60-1502 60-1510 |
Smiths Medical has identified that the Duralife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the Duralife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.
Recall start date: March 30, 2023