· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Reliance Vision Multi-Chamber Washer Disinfector | FH92-072 FH91-072 FH82-072 FH81-072 FH82-042 |
Steris has identified that in the remote occurrence in which the electrical contactor component present in the drying chambers of the reliance vision multi-chamber washer/disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Recall start date: March 29, 2023