· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Trilogy EV300 | CA2200X12B |
| All lots. | Trilogy EVO | CA2110X12B |
Philips Respironics has detected environmental contamination (e.g. external dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on a flow sensor within the device. This may cause the device to inaccurately deliver pressure or volume/flow.
Recall start date: March 28, 2023