Health Products

Philips Incisive CT

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Philips Incisive CT
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Incisive Ct728143

Issue

Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover

The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces).

Recall start date: May 1, 2023