Health Products

Horizon® Microclip™ Titanium Ligating Clips with Tape

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Horizon® Microclip™ Titanium Ligating Clips with Tape
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Horizon® Microclip™ Titanium Ligating Clips with Tape005200

Issue

Teleflex is initiating this voluntary recall because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures. Additionally, this introduces a further risk to the patient whereby the tips of the applier may project further past the tissue to be ligated, potentially risking injury to other distal anatomical structures.

Recall start date: May 2, 2023