Health Products

Single Use Repositionable Clip

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Single Use Repositionable Clip
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Single Use Repositionable ClipHX-202LR
HX-202UR

Issue

Olympus has become aware of increased complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

The issues described above can lead to hazardous situations in which the clips cannot achieve their intended purpose, namely: (1) approximate a tissue opening or defect, (2) provide hemostasis to a bleeding vessel or mucosa, or (3) serve as a marker for subsequent location identification. Patient harms associated with these hazardous situations include: absence of treatment (e.g. hemostasis not achieved), prolonged procedure (e.g. extended time to accomplish hemostasis), and unexpected medical interventions (e.g. additional clipping) being required to accomplish therapeutic purposes. Additionally, the issues described above can lead to mucosal injury/perforation, hemorrhage, and dehiscence of a previously approximated closure.

Recall start date: May 2, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime