Health ProductsType I recall

Radiant Warmer and Birthing Room Warmer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Radiant Warmer and Birthing Room Warmer
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
RW82-1C
RW82VHA-1
RW82VHA-1C
RW82VHA-1D
RW82-1
RW82-1D
Radiant WarmerMU20503
WBR82-1
WBR82-1C
Birthing Room WarmerMU20512

Issue

During the course of global market surveillance activities, the manufaturer has become aware of cases in which the resuscitaire infant radiant warmer with the optional scale could display inaccurate weight values. There is a potential for the mattress cover material to slightly interfere with the corner brackets and/or side panels of the bassinet. When this condition is present, the scale may not "zero" correctly during a "weigh" function, which may lead to an inaccurate weight measurement of the infant.

Recall start date: April 28, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime