Health Products

Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Physical medicine
Product: Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
1.0.02
1.0.04
1.0.01
1.0.03
1.0.00
Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist DevicePUSHTRACKER E3
PUSHTRACKER E2

Issue

Through design and development work, Max Mobility has identified a software issue with the application. When multiple processes are running on the watch's central processing unit, the application may crash unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Recall start date: May 2, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime