Health Products

Vitros Chemistry Products Calibrator Kit 20

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Vitros Chemistry Products Calibrator Kit 20
Issue: Medical devices - Performance
What to do: Review/post notice (forward if product is distributed), discontinue/discard inventory, return reply form.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
2022Vitros Chemistry Products Calibrator Kit 20680 1704

Issue

The Vitros Chemistry Products Calibrator Kit 20 is used to calibrate Microtip assays C3, C4, IgM, IgA, IgG, and transferrin. Vitros Calibrator Kit 20 contains five calibrator levels. Although lot 2022 met all initial release criteria, vial to vial variability was subsequently observed during routine internal testing. This prompted an investigation that confirmed that the level 5 calibrator exhibited vial to vial variability which may cause a bias for all six affected assays, although to varying magnitudes.

Recall start date: May 9, 2023