Health Products

Philips Patient Information Center

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Philips Patient Information Center
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Patient Information CenterVERSION 4.0

Issue

During extended operation of the MX40 with the PIC iX, the DHCP (dynamic host configuration protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the pic ix, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

Recall start date: May 12, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime