Alert type: Health product recall
Category: Health products - Medical devices
Product: Platelia Aspergillus IgG
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
2L0039Platelia Aspergillus IgG62783

Issue

Bio-Rad Laboratories is initiating a recall due to a confirmed quality anomaly for Platelia aspergillus IgG test (cat #62783), for lot 2l0039 (expiry date: 2024/04/15). The current lot 2l0039 shows a decrease of Optical Density (OD) of all calibrators. The OD of the R4A calibrator 10 au/ml could be lower than 0.200 (the expected quality criteria is R4A OD value > 0.200) that could result in an invalid run.