Health ProductsType I recall

Oxylog 3000 Plus - Basic Unit

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Oxylog 3000 Plus - Basic Unit
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Oxylog 3000 Plus - Basic Unit5704811

Issue

During the course of global market surveillance activities, the manufacturer has become aware of cases where Oxylog 3000 Plus emergency and transport ventilators have stopped ventilation due to empty batteries despite being reconnected to the mains power supply after previous battery operation. This is due to a timing problem in the firmware of the charging circuit PCB, where the charging process may be interrupted, making the switch to mains operation impossible.

Recall start date: May 17, 2023

Get told about the next health product recall.
Get health product recall alerts