Health Products

Biofire® Respiratory Panel 2.1 (Rp2.1)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Pathology
Product: Biofire® Respiratory Panel 2.1 (Rp2.1)
Issue: Medical devices - Performance
What to do: Contact the importer, Biomerieux Canada Inc.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
2168222Biofire® Respiratory Panel 2.1 (Rp2.1)423742

Issue

Biomerieux identified a manufacturing event that may contribute to elevated rates of false negative results while using the Biofire RP2.1 panel. Biomerieux has not received any false negative complaints from the field but, out of an abundance of caution, Biomerieux is recalling the product that was manufactured during this event.

Recall start date: May 16, 2023