Health Products

Ultraview SL Command Module

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Ultraview SL Command Module
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Pre-Software Version 2.04.00
Post Software Version 2.04.00
Ultraview SL Command Module91496

Issue

Software versions prior to 2.04 show a waveform while in suspend processing, while software version 2.04 and above do not show a waveform. Both software versions show an error message stating "ECG processing suspended" in the waveform zone, "?" in the heart rate, SpO2, and ST segment values, and a message stating "ECG ALM off" will replace the alarm limits. Hospitals may have multiple products concurrently running multiple software versions. The causal chain would be the user who is unaware of the different versions failing to recognize suspend state due to the presence of the waveform, leaving the patient in suspend with alarms inactive, a cardiac event occurring, and user failing to detect the condition with alarms inactive.

Recall start date: May 23, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime