Health Products

MobileDiagnost wDR

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: MobileDiagnost wDR
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.MobileDiagnost wDR9890 010 8952X

Issue

Philips has become aware of a software login issue associated with specific models of Mobilediagnost wDR systems running on Windows 10 operating system preventing use of the system. This issue occurred because the software certificate protecting the Eleva user login expired on May 15, 2023. Therefore, if the customer logs out or reboots the system, they will be prevented from logging in. The issue is in regard to expiry of 6 certificates of Eleva software by 14th May 2023. Eleva systems is unable to function after these dates. Due to the Eleva certificate expiry, the system will not be able to log in and therefore rendering the system non-usable.

Recall start date: May 31, 2023