Health ProductsType I recall

Dose IQ Safety Software

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Dose IQ Safety Software
Issue: Medical devices - Performance
What to do: Contact manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
All lots.DOSEIQW0001Dose IQ Safety Software

Issue

Baxter Corporation is issuing an urgent medical device correction due to a defect in the Dose IQ Safety Software used with Novum IQ Large Volume Pump, resulting in an invalid initial setting for the air-in-line threshold for any new care area created in which none of the settings are changed.

Recall start date: May 29, 2023

Get told about the next health product recall.
Get health product recall alerts