· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | 650705550001 | 6507 Power Pro 2, High Config |
| 2211003786 2211003736 2211003737 2211003725 2211003726 2211003742 2211003768 2211003734 2211003752 2211003763 | 650705550002 | 6507 Power Pro 2, Mid Config |
Stryker has identified that the base leg assembly may bend. It was discovered that the variability in the manufacturing process at a tier 2 supplier may have results in low tensile strength of material. Bent base leg assembly may result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.
Recall start date: May 26, 2023