Health Products

Giraffe Omnibed Carestation Cs1

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Giraffe Omnibed Carestation Cs1
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.2082844-001Giraffe Omnibed Carestation Cs1

Issue

There is the potential that an incorrect type of secondary latch could have been installed on the northeast (ne) and/or northwest (nw) bedside panels for certain Giraffe Omnibed Carestation devices (serial number specific). With the incorrect type of secondary latch installed, the north wall can potentially disengage when canopy is raised. However, the device is designed so the panels will stay closed and function as intended, even if only the primary latches on the south end are properly engaged.

Recall start date: June 2, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime