Health Products

Datascope Cardiosave Hybrid And Rescue Intra-Aortic Balloon Pumps (IABP)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Datascope Cardiosave Hybrid And Rescue Intra-Aortic Balloon Pumps (IABP)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Datascope Cardiosave Hybrid And Rescue Intra-Aortic Balloon Pumps (IABP)0998-00-0800-XX

Issue

1: There have been reports of IABP units losing the ability to charge batteries in one or both bay slots due to a failure in the power management board charging path circuitry due to damage to the component from a previous electrical surge. Therapy may be interrupted if batteries fail to charge and the device is disconnected from ac power. Low battery alarms may alert the user to the issue prior to interruption of therapy.

Issue 2: An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the power management board and/or solenoid board. This issue may lead to an unexpected interruption of therapy.

Recall start date: June 9, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime